Tech insight · Tubes and sterility
What "gamma-sterile" means for a blood collection tube
How gamma irradiation sterilises evacuated blood tubes, what a sterility assurance level of 10⁻⁶ means, and the evidence to ask any tube supplier for.
Many evacuated blood collection tubes are labelled "sterile interior". Ours are sterilised with gamma radiation, and so are many other manufacturers' tubes. Gamma sterility is common in the industry; what differs is the evidence behind it. This article explains what the claim means and what a laboratory can check.
Why a sterile interior matters
A tube's interior is in contact with the patient's blood through the needle. In rare circumstances, if the tube is above the arm and the tourniquet is released, blood can flow back from the tube towards the vein. A sterile interior removes the theoretical risk of introducing microorganisms that way. It also protects samples for tests where microbial contamination could interfere with results.
What sterile means: the SAL
Sterilisation kills microorganisms in a predictable, exponential way. It is not possible to prove that every single unit is sterile, so sterility is expressed as a probability: the sterility assurance level.
European standard EN 556-1 allows a terminally sterilised device to be labelled STERILE only when the SAL is 10⁻⁶ or better, meaning a probability of no more than one in a million that a viable microorganism survives on a unit.
How gamma sterilisation works
Gamma rays from a cobalt-60 source pass through the boxed product. They damage the DNA of microorganisms so they cannot reproduce. Because gamma penetrates the closed cap, the evacuated tube and the outer cartons, the product is sterilised in its final, sealed state.
Key features:
- Controlled by one parameter, dose, measured in kilograys (kGy). Temperature, humidity and gas concentration do not have to be controlled as they do for ethylene oxide.
- No residues. Ethylene oxide sterilisation requires residue limits under ISO 10993-7; gamma leaves nothing to outgas.
- Material compatibility must be validated. Radiation can affect some plastics and additives, so the manufacturer must show that the vacuum, additive and gel still perform after the maximum dose.
The standards behind the claim
| Standard | What it covers |
|---|---|
| ISO 11137-1 | Requirements for developing, validating and routinely controlling radiation sterilisation |
| ISO 11137-2 | Establishing the sterilisation dose, for example by the VDmax method for a dose of 25 kGy |
| ISO 11137-3 | Dosimetry: measuring the dose delivered |
| EN 556-1 | The SAL of 10⁻⁶ required for a STERILE label |
| ISO 6710 | Single-use containers for human venous blood specimen collection |
Validation and routine control
A manufacturer validates gamma sterilisation in three parts:
- Bioburden: measure how many microorganisms are on the product before sterilisation.
- Dose setting: show by testing that a chosen dose achieves an SAL of 10⁻⁶ for that bioburden.
- Dose mapping: map the dose throughout a loaded carton, so the minimum dose is at least the sterilisation dose and the maximum dose does not harm the product.
Then, for every batch, dosimeters placed with the product confirm the dose delivered. This is called dosimetric release. Periodic dose audits, typically quarterly, confirm the sterilisation dose is still valid as bioburden changes.
Questions to ask any tube supplier
- Which validation method do you use under ISO 11137-2?
- What is the minimum and maximum dose, and how is the product shown to perform at the maximum?
- Is every lot released on dosimetry records, and can I see the certificate?
- How often are dose audits done, and when was the last one?
- Is product performance (vacuum, draw volume, additive) tested after sterilisation?
For DBO tubes, the Certificate of Quality for each lot records the sterility status alongside visual, physical, chemical and microbial inspection. You can look it up by lot number.
References
- ISO 11137-1:2006 (amended). Sterilization of health care products: Radiation. Part 1: Requirements for development, validation and routine control. International Organization for Standardization.
- ISO 11137-2:2013. Sterilization of health care products: Radiation. Part 2: Establishing the sterilization dose. International Organization for Standardization.
- EN 556-1:2001. Sterilization of medical devices: Requirements for medical devices to be designated "STERILE". Part 1: Requirements for terminally sterilized medical devices. CEN.
- ISO 6710:2017. Single-use containers for human venous blood specimen collection. International Organization for Standardization.
- CLSI. GP41: Collection of Diagnostic Venous Blood Specimens. 7th ed. Clinical and Laboratory Standards Institute; 2017.