Gel separator tubes on analysers: fibrin, clots, gel and the sample probe

Why serum gel tubes cause probe clogs, short samples and level-detection errors, and the clotting, centrifuge and loading steps that prevent them.

By DBO Technical and Scientific TeamChecked against the guidelines cited below on 25 October 2026Published 25 October 2026 · 3 min read
WhatsAppLinkedInEmail

Gel separator tubes are designed so that, after centrifugation, a polymer gel moves up and settles between the cells and the serum or plasma. The analyser probe then samples from above the gel. When any step goes wrong, the result is a probe clog, a "sample short" flag, a level-detection error or, at worst, a wrong result with no flag at all.

Fibrin: the main culprit

Serum tubes contain a clot activator, usually silica particles, that speeds clotting. If the tube is spun before clotting is complete, the serum contains fibrinogen that continues to clot after centrifugation, forming strands or gel-like masses. These are almost invisible and are drawn into the probe.

Prevent it:

  • Invert the tube gently straight after collection, as the IFU specifies, so the activator mixes with all the blood.
  • Let it clot upright for the full time in the IFU, commonly 30 minutes at room temperature.
  • Allow longer for patients on heparin or other anticoagulants, or use a heparin plasma tube instead, if the test allows.
  • Do not refrigerate before clotting; cold slows clotting.

The gel barrier

The gel must move cleanly to the boundary between cells and serum. Problems include:

  • Incomplete barrier, from too low a force, too short a spin or a fixed-angle rotor. The probe may reach cells.
  • Slanted barrier, typical of fixed-angle rotors. Use swing-out rotors unless the tube IFU allows fixed-angle.
  • Gel strands or droplets in the serum, sometimes from very high force or heating in the centrifuge. They can coat the probe.
  • Re-centrifugation. Spinning a gel tube a second time disturbs the barrier and lets potassium, phosphate and LDH leak from cells.

Level detection and surface problems

Most analysers find the liquid surface by measuring capacitance or pressure as the probe descends. Anything at the surface that is not serum confuses it:

  • Bubbles and foam, from vigorous mixing, pouring or uncapping roughly.
  • Fibrin films across the surface.
  • Condensation or droplets inside the cap, from cold storage.

Pop bubbles with a clean applicator stick or let the sample stand. Never shake a tube to "mix" it before loading.

Fit on the analyser

Tube diameter, length, cap type and barcode position must match the analyser's rack and reader. A tube that sits too high can collide with the probe; one that sits too low can give "sample short". A barcode label applied over the manufacturer's label, or crooked, may not scan. DBO's instrument compatibility tool lists tube sizes against more than 240 analyser models, with a troubleshooter for these errors.

A short checklist for the lab

  1. Was the tube filled to its nominal volume and inverted?
  2. Did it clot upright for the full time?
  3. Was it spun at the IFU × g and time, in a swing-out rotor?
  4. Is the barrier complete and level?
  5. Is the surface free of bubbles, foam and films?
  6. Was it spun once only?

References

  1. CLSI. GP44: Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests. 4th ed. Clinical and Laboratory Standards Institute; 2010.
  2. Simundic AM, Bölenius K, Cadamuro J, et al. Joint EFLM-COLABIOCLI Recommendation for venous blood sampling. Clin Chem Lab Med. 2018;56(12):2015–2038. PubMed
  3. ISO 6710:2017. Single-use containers for human venous blood specimen collection. International Organization for Standardization.
Free DBO toolCheck your analyser and tube fit

Read next

New articles by email

One short email when we publish something new. No other mailings. Unsubscribe with one click.