Convert centrifuge rpm to relative centrifugal force (× g) and back, measure rotor radius correctly, and set spin conditions for serum, plasma and gel tubes.
Tube manufacturers state spin conditions as a force, for example "1,300 to 2,000 × g for 10 minutes", because force is what separates cells from plasma and forms the gel barrier. Most centrifuges, however, are set in revolutions per minute. Converting between them is simple, but it depends on one number many labs get wrong: the rotor radius.
The calculator
The formula
Relative centrifugal force (RCF), expressed as multiples of the Earth's gravity (× g), is:
RCF = 1.118 × 10⁻⁵ × r × rpm²
and the reverse:
rpm = √( RCF ÷ (1.118 × 10⁻⁵ × r) )
where r is the radius in centimetres. The constant 1.118 × 10⁻⁵ comes from converting revolutions per minute into angular velocity and dividing by the acceleration of gravity.
Because rpm is squared, small changes have big effects. Doubling the rpm quadruples the force.
Measuring the radius
The radius that matters is the maximum radius: the distance from the centre of the rotor shaft to the bottom of the tube while it is spinning.
- Swing-out rotor: measure with the bucket swung out horizontally, to the bottom of the bucket where the tube base sits.
- Fixed-angle rotor: measure from the centre to the bottom of the tube cavity along the radius.
Most centrifuge manuals list r(max) for each rotor and adapter. Use that figure if you have it. Adapters for small tubes can change the radius, so check the figure for the adapter you use.
Example
A tube IFU asks for 1,500 × g. Centrifuge A has r = 15 cm; centrifuge B has r = 10 cm.
- A: rpm = √(1,500 ÷ (1.118 × 10⁻⁵ × 15)) ≈ 2,990 rpm
- B: rpm = √(1,500 ÷ (1.118 × 10⁻⁵ × 10)) ≈ 3,660 rpm
Setting both to 3,000 rpm would give about 1,510 × g on A but only about 1,010 × g on B: too little to form a reliable gel barrier.
Getting the spin right for blood tubes
- Serum tubes must clot first. Allow the clotting time in the IFU, commonly 30 minutes upright at room temperature, longer for patients on anticoagulants. Spinning too early leaves fibrin.
- Follow the IFU for force and time. Typical ranges for serum, heparin and gel tubes are 1,300 to 2,000 × g for 10 minutes. Coagulation testing needs platelet-poor plasma and often a longer spin; follow your coagulation method's requirement.
- Prefer swing-out rotors for gel tubes. They form a flat, even barrier. Fixed-angle rotors give a slanted barrier; use them only where the tube IFU allows.
- Never re-spin a gel tube after the barrier has formed. Potassium and other analytes leak from the cells into the serum.
- Balance the load with tubes of equal fill.
- Check the centrifuge at least yearly with a calibrated tachometer and timer, and record it.