Haemolysis at collection: the common causes and how to cut your rate

Haemolysis is the top reason samples are rejected. The collection and handling causes, which results it changes, and a plan to measure and cut your rate.

By DBO Technical and Scientific TeamChecked against the guidelines cited below on 25 October 2026Published 25 October 2026 · 4 min read
WhatsAppLinkedInEmail

Haemolysis, the rupture of red cells with release of their contents into serum or plasma, is the most common reason a laboratory rejects or flags a sample. In emergency departments the rate can be several times higher than in outpatient clinics. Each haemolysed sample means a repeat needle for the patient and a delay in treatment.

The good news is that almost all of it is created after the blood leaves the vein, which means it can be prevented.

In vivo or in vitro?

In vivo haemolysis happens inside the patient: haemolytic anaemias, transfusion reactions, mechanical heart valves, severe infections. It is a clinical finding and must be reported, not rejected.

In vitro haemolysis happens during or after collection. It accounts for the great majority of haemolysed samples, and it is the laboratory's and the collector's problem to fix.

If several consecutive samples from a patient are haemolysed despite good technique, tell the laboratory. It may be in vivo, and that matters for the patient.

Collection causes

CauseWhy it haemolysesWhat to do
Drawing through an IV catheterTurbulent flow through narrow plastic, often with a syringeDraw by venepuncture where possible. If a line must be used, follow your local line-draw procedure
Difficult or prolonged drawRepeated probing, slow trickling flowChoose the vein carefully; if it fails, stop and re-site
Very small needleHigh shear as blood is pulled throughUse 21G for routine adults; 22G or 23G for small, fragile veins
Syringe transferForcing blood through the needle or pulling the plunger hardUse a safety transfer device; let the vacuum draw the blood
Wet alcoholAlcohol entering the sampleLet the site dry fully
Tourniquet too longStasis and fragile cellsOne minute maximum; release when blood flows
Underfilled tubeExcess vacuum pulls blood through hard; wrong additive ratioLet each tube fill until the vacuum is exhausted
Shaking the tubeMechanical damageGentle, full inversions only

Handling and transport causes

  • Pneumatic tube systems with high speeds and sharp turns. Ask your engineers for acceleration data and test with paired samples.
  • Temperature extremes: freezing whole blood, or leaving tubes on a hot dashboard.
  • Delayed separation of serum or plasma from cells, especially in the heat.
  • Re-centrifuging gel tubes after the barrier has formed.
  • Centrifuge settings far above the IFU, or a failing brake.

Which results change

Red cells contain far more potassium, LDH, AST and phosphate than plasma, so even slight haemolysis raises them. Free haemoglobin also interferes optically with many photometric methods, and enzymes released from cells interfere with some immunoassays. Insulin and troponin results can also be affected depending on the method.

Each method has its own tolerance. The analyser's haemolysis index, together with the manufacturer's interference limits, is the only reliable way to decide whether a result can be released. Visual inspection misses low levels and over-calls others.

A practical reduction plan

  1. Measure. Use the haemolysis index on every chemistry sample. Calculate the rate as haemolysed samples per total samples, monthly, by location and, where policy allows, by collector.
  2. Find the hot spots. In most hospitals one or two areas, usually the emergency department, produce most of the haemolysis.
  3. Fix the cause in that area. Line draws, syringe use and needle size are the usual levers.
  4. Train and observe. A short observed draw with feedback changes practice more than a lecture.
  5. Compare with peers. The IFCC Working Group on Laboratory Errors and Patient Safety runs a quality indicator programme with published performance specifications. DBO's free anonymous pre-analytical benchmark lets you compare your haemolysis rate with laboratories of the same size and type.

References

  1. Lippi G, Plebani M, Di Somma S, Cervellin G. Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories. Crit Rev Clin Lab Sci. 2011;48(3):143–153.
  2. Simundic AM, Bölenius K, Cadamuro J, et al. Joint EFLM-COLABIOCLI Recommendation for venous blood sampling. Clin Chem Lab Med. 2018;56(12):2015–2038. PubMed
  3. Sciacovelli L, Padoan A, Aita A, Basso D, Plebani M. Quality indicators in laboratory medicine: state-of-the-art, quality specifications and future strategies. Clin Chem Lab Med. 2023;61(4):688–695.
  4. CLSI. GP41: Collection of Diagnostic Venous Blood Specimens. 7th ed. Clinical and Laboratory Standards Institute; 2017.