Tech insight · Processing
Glucose samples: why fluoride alone is not enough
Glucose falls about 5 to 7% an hour in a blood sample. Why fluoride works too slowly, and the two options the 2023 ADA and NACB guidelines recommend.
A glucose result is only as good as the time between the needle and the centrifuge. Red and white cells continue to use glucose after collection, so the result falls steadily until the cells are removed or their metabolism is stopped.
How fast glucose falls
At room temperature, glucose in whole blood decreases by about 5% to 7% per hour. The loss is faster with high white cell counts and at higher temperatures. In a sample that waits two hours before it is spun, a true glucose of 7.0 mmol/L (126 mg/dL), the diagnostic threshold for diabetes, can read below the threshold.
Why fluoride is not enough
Grey-top tubes contain sodium fluoride, often with potassium oxalate. Fluoride inhibits enolase, an enzyme late in glycolysis. But the earlier steps continue, and glucose is still consumed until the pathway backs up. In practice, fluoride does not fully stabilise glucose for the first one to several hours after collection. Its value is in stopping the loss after that point.
What the guidelines recommend
The 2023 guidelines from the American Diabetes Association and the National Academy of Clinical Biochemistry (now part of ADLM), published by Sacks and colleagues, recommend one of two approaches for accurate plasma glucose:
- Tubes containing a citrate buffer, usually with fluoride. Acidifying the sample inhibits the early glycolytic enzymes immediately, so glucose is stable from the moment of collection.
- If such tubes are not available: place the tube immediately in an ice-water slurry and centrifuge to remove the cells within 15 to 30 minutes.
Collecting into a fluoride tube and leaving it in a rack at room temperature is not one of the options.
Practical points
- Know which glucose tube your lab uses and its IFU. Citrate-buffered tubes may have a slightly different fill and mixing requirement.
- Draw the glycolytic inhibitor tube last in the order of draw.
- Mix fully so the inhibitor reaches all the cells.
- For oral glucose tolerance tests and pregnancy screening, sample handling matters most, because the thresholds are close together.
- Point-of-care glucose meters use whole blood immediately and are not affected by this delay, but they are not a substitute for laboratory diagnosis.
References
- Sacks DB, Arnold M, Bakris GL, et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem. 2023;69(8):808–868. PubMed
- CAP Today. New guidelines published on lab analysis in diabetes. 2023. CAP Today
- CLSI. GP44: Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests. 4th ed. Clinical and Laboratory Standards Institute; 2010.